D14 HZ Module
Housed microporous polypropylene hollow fiber membrane degassing / contactor module. Degassing water/liquid in benchtop systems, laboratory analyzers and OEM devices.
-
Polypropylene hollow fiber membrane
-
232cm² P60 active membrane surface
-
Transparent PSU housing
-
Luer Lock, hose barb D5mm port connections



Characteristic
Membrane Type | P60 | P80 | |
Membrane | Material | Polypropylene (natively hydrophobic) | |
Pore size | average 0.1 μm | ||
OD/ID | 310/240 μm | 530/440 μm | |
Burst / Collapse pressure | >5.5 / >3.5 bar | ||
Module | Active Membrane Surface | 232 cm² | 256 cm² |
Priming Volume Shellside | ml | ml | |
Priming Volume Lumenside | ml | ml | |
Housing Material | PSU transparent | ||
Potting Material / Seals | Polyurethane / glue | ||
Shell pressure | < 5 bar | ||
Temperature | < 45°C | ||
Ports | Luer Lock, hose barb D5mm | ||
Dimensions | D20 x 127mm | ||
Weight | 18g |
① food grade, no additives PP is used, but due to production small amounts of other materials might be used, please consult us for chemical resistance
② caps are glued on
③ lumen side - luer locks, shell side hose barb D5mm

Possible Applications
- Gas transfer, microporous hollow fiber membrane is used so gas/liquid or liquid/gas
- Gas filtration - gas/solids separation on microfiltration level
- Membrane degassing - lab, OEM applications
Applications Notes
- Shell side ports are hose barb, usually vacuum is applied. Use PU or thick wall PVC hose.
- Try to avoid temperature shocks, slow change is preferable. For contactor (liquid / gas, degassing) applications temperature gradients might cause condensation inside the pores, wetting and module bleeding. When this happens thorough drying (dry air @50°C at least 24 hours) will restore the module.
- No oxidizing chemistry or UV light – PP membrane is very sensitive. For chemical cleaning use standard NaOH / citric acid procedure.
- Chemical resistance can be estimated from used materials PP, PSU, polyurethane, usually least resistant is potting - polyurethane.
Custom Options
There are other custom variations of the D14 HZ Module:
- Membrane P80, custom or customer supplied
- Solvent resistant glue for caps
Dimensions and Options